GEORGE PHILIPS MEDICAL ENGINEERING PVT. LTD.
 
Home » Products » Physiological Monitoring System » Disposable Pressure Transducer

Disposable Pressure Transducer



<< Previous   Disposable Pressure Transducer

Product Code : 25

Product Specification

Disposable Pressure Transducer

FEATURES

BENEFITS

Modern mems chip technology

Accurate transmission of pressure waves

Transparent housing

For easy detection of air bubbles and quick priming

Automatic continuous flow @
3 cc/hr or 30 cc/hr

Maintains catheter patency and eliminates the need for intermittent manual flushing.
Facilitates dynamic response testing of the monitoring system, allowing rapid flushing of Pressure Monitoring Line after collection of blood sample.

Assembled & individually tested with automated machines.

High quality, safety and reliability of each unit.

Multiple cable end options

Compatible with most of the commonly used patient monitors.


MATERIAL SPECIFICATIONS

Body & end adaptors

Polycarbonate

Polycarbonate

Silicone

Capillary flow system

Glass

Luer Caps

Polyethylene

TECHNICAL SPECIFICATIONS

Pressure Stability

5 Bar Hydrostatic

Priming Volume

0.80 ml

Continuous flow rate at 300 mm Hg pressure

3 cc / hr

Activated flush rate at 300 mm Hg pressure

130 ml / min

PRODUCT CODE & PACK DETAILS

GP/HM/03/XXX

For detailed specifications & product codes, kindly refer table at the end of the catalogue



Specifications

Parameter

Value

Units

Notes

General

Pressure Ranges

– 30 to 300

mmHg

–0.58 psi to 5.8 psi (-0.03 bar to 0.03 bar)

Overpressure

125

psi

(x bar) minimum

Electrical @ 72°F (22°C) Unless Otherwise Stated

Input Excitation

1 to 10

VDC

Calibrated for 6 VDC

Dielectric Breakdown

10,000

VDC

5

Risk Current

2

µA

Maximum (per AAMI), 5

Input Impedance

1800 to 3300

Ω


Output Impedance

285 to 315

Ω


Environmental

Temperature

Compensated

59° to 104°

°F

(15°C to 40°C)

Operating Storage

59°F to 104°F

°F

(15°C to 40°C)

–13° to 158°

°F

(–25°C to 70°C)

Humidity

10 to 90

%


Light Sensitivity

1

mmHg

maximum (per AMMI BP22)

Operating Product Life

168

hours


Shelf Life

3

years


Mechanical

Weight

< 0.0044

lb

(<2 g)

Volume Displacement

< 0.0008

in

(<0.02 mm3)

Media Interface

Medical grade, dielectric gel

Gel Tube Interface Material

Polycarbonate



Performance Parameters *

Units

Minimum

Type

Maximum

Notes

Offset

mmHg

– 25

0

25


Sensitivity

V / V / mmHg

4.95

5

5.05


Calibration

mmHg

97.50

100

102.50

2

Symmetry

%

—

—

± 5


Linearity(–30 to 100 mmHg)

mmHg

—

—

1

6

Linearity (100 to 200 mmHg)

% output

—

—

1

6

Linearity (200 to 300 mmHg)

% output

—

—

1.50

6

Thermal Coefficient Offset

mmHg / °C

—

—

± 0.30

3

Thermal Coefficient Span

% / °C

—

—

± 0.10

3

Frequency Response

Hz

1200

—

—

5

Phase Shift

degrees

—

—

5

5

Offset Drift

mmHg/8hrs

—

—

1

4.50

Other Products in 'Physiological Monitoring System' category
CVP Manometer Set
CVP Manometer Set
Pressure Monitoring Line With Transparent Tube
Pressure Monitoring Line With Transparent Tube
Continuous Flush Or Flow Device
Continuous Flush Or Flow Device
Spinal Manometer Set
Spinal Manometer Set
Pressure Monitoring Kit
Pressure Monitoring Kit
 
<< Back


 
 
 


  Contact Details
GEORGE PHILIPS MEDICAL ENGINEERING PVT. LTD.
304, 3rd Floor, Gemstar Commercial Complex, Ram Chandra Lane Extension, Kachpada, Malad, Mumbai - 400064, Maharashtra, India
Phone:91-22-28442884/28442885
Fax:91-22-28442885/28769969
Send Mail   Send SMS   
Key Personnel
Dr. Vinita Murthy (Director - Exports)
Mobile:+919223113665
Mr. P. R. Joshi (Vice President - Sales)
Mobile:+919930920453









Copyright © 2011-12 by GEORGE PHILIPS MEDICAL ENGINEERING PVT. LTD. All Rights Reserved.